TEST EVALUATION
TMS PACK
vs. Laboratory Screening & GC/MS

EVALUATION OVERVIEW
This evaluation was conducted to compare the accuracy of the TMS PACK Drugs of Abuse Test Kits versus traditional laboratory screening methods. The evaluation was performed at a SAMSHA certified laboratory. 194 samples were initially screened by the laboratory using a HITACHI Analyzer, (DRI Liquid Reagents). These samples were then tested using the TMS PACK test kits. Positive samples, either HITACHI or TMS PACK were then tested using GC/MS. (Note: If the laboratory screen result and TMS PACK indicated a negative result, no further testing was conducted.)

TMS PACK TEST KITS

TOTAL SAMPLES 194
TOTAL CORRECT SAMPLES
OVERALL ACCURACY
185
95.5%
TOTAL INCORRECT SAMPLES 9
TOTAL FALSE NEGATIVE SAMPLES
FALSE NEGATIVE RATE
0
0%
TOTAL FALSE POSITIVE SAMPLES
FALSE POSITIVE RATE
9**
4.5%

TMS PACK reported one sample to contain THC/COC. GC/MS determined that the sample contained 8 ngl/ml THC & 104 ng/ml COC. Five samples were reported as positive for Methamphetamine or Amphetamine, one sample was reported as positive for Morphine and Methamphetamine and one sample was reported as positive for Morphine. GC/MS determined these samples to be negative. The positive results for Methamphetamine/Amphetamine are believed to be the result of cross-reactivity with OTC products.

LABORATORY SCREENING

TOTAL SAMPLES 194
TOTAL CORRECT SAMPLES
OVERALL ACCURACY
187
96.0%
TOTAL INCORRECT SAMPLES 7
TOTAL FALSE NEGATIVE SAMPLES
FALSE NEGATIVE RATE
3**
1.5%
TOTAL FALSE POSITIVE SAMPLES
FALSE POSITIVE RATE
4**
2.0%

Three samples screening negative for THC confirmed by GC/MS for THC. Four samples reported as positive for Methamphetamine or Amphetamine were negative when tested by GC/MS.

NOTE Screening results were adjusted to represent a Morphine cut-off level of 300 ngl/ml.